M600 Helium Compressor Specifications
Maximum Helium Flow 60.0 CFM
Static Helium Charge 240 psi (+/- 5 psi)
Water Cooled
Minimum Required Water Flow 1.5 gallons /minute (5.7 liters/minute)
Water Temperature Operating Range 60 – 80° F (16 – 27° C)
Cooling Water Quality Municipal Drinking Water or Better
Dimensions (height X width X depth) 21” X 17.25” X 19.75” (53.3 X 43.8 X 50.2 cm)
Weight 260 lbs. (118.2 kg)
Air Cooled
Air Temperature Operating Range 65 – 105° F (18 – 41° C)
Dimensions (height X width X depth) 26” X 17.25” X 32” (66 X 43.8 X 81.3 cm)
Weight 300 lbs. (136.1 kg)
Power
Low Voltage Input 190 – 260 VAC
High Voltage Input 345 – 525 VAC
Consumption 9.0 kilowatts
3 Phase 50 or 60 Hz
Recommended Breaker Rating 30 Amp